SD-508-ATP-B Environmental Testing Chamber
Illustration of man next to  for scale
 Environmental Testing Chamber
Illustration of man next to  for scale

SD-508-ATP-B

BenchtopRack and Stack

TUVRheinland icon Benchtop Stackable icon

The SD-508-ATP-B is a benchtop-style environmental test chamber for temperature-only testing.  The chamber is designed to fit on your lab table and still have easy and safe access when you reach-in the workspace.  

If you are concerned about losing your lab counter space, this test chamber has an optional floor stand for one chamber or get double the test workspace when you stack two SD Test Chambers in the same footprint.

STRUCTURE.  The internal workspace is lined with type 304 stainless steel for a vapor-tight internal construction. Two specialized ports at the back of each test chamber allow for cable patches to feed through the back of the test chamber workspace and easily connect to your battery cycler test equipment.

The exterior has a durable powder-coat finish over heavy-gauge steel for a long-lasting clean look. Between the two layers is highly-efficient, low k-factor, thermal insulation. This chamber is built to last for many years.

CIRCULATION SYSTEM.  Air circulation is crucial for environmental testing. The SD-508-ATP-B Test Chamber has a proprietary baffle and fan design combination. This optimizes temperature conditions within the workspace for the ATPPRIME battery fixture. 

HEATING SYSTEM.  SD test chambers feature heating systems equipped with open ceramic-core nichrome heaters for reliable and accurate temperature control within ±0.5°C.

COOLING SYSTEM.  The cooling system in the SD-508-ATP-B consists of a self-contained mechanical refrigeration system with a single stage configuration, and an air-cooling condenser. An optional water cooling-condenser is available.  

Add boost cooling with liquid LN2 or CO2 for even faster pull-down rates. It's a great option for stress tests that require rapid cooling.  

Even without boost cooling, AES’ standard cascade refrigeration systems have the fastest standard change rates of any environmental test chamber manufacturer. See Basic Specifications tab below. 

INSTRUMENTATION. SD-508-ATP-B Test Chamber incorporates AESONE CONNECT creating a robust connected device giving you a standard and uniform interface that makes operation easier. Learn more in the Instrumentation Tab below.

ALTERNATIVE TEMPERATURE RANGES:

  • SD-208 -70°C to +180°C (-94°F to +356°F)
Buy $39,924 (base price)
Show More...
  • Specifications
  • Standard Features
  • Instrumentation
  • Safety Features
  • Optional Features
  • Downloads

Specifications

SD Series



Illustration of a 6-foot tall man
Height: 72" (6' tall)

SD-202 SD-502

SD-202
Height: 25.5"
Volume: 2.083 cu. ft.

SD-208 SD-508 SD-508-ATP

SD-208
Height: 35"
Volume: 8 cu. ft.

SD-508-SAFE

SD-508-SAFE
Height: 35"
Volume: 8 cu. ft.

SD-501 SD-501-ATP SD-501-SAFE

SD-501
Height: 16.73"
Volume: 1.01 cu. ft.

SD-508-ATP-B

SD-508-ATP-B
Height: 81.3"
Volume: 16 cu. ft.

SD-501-3RS

SD-501-3RS
Height: 78.5"
Volume: 3.03 cu. ft.
Interior: 24.0W x 24.0D x 24.0H (in)
  609.5W x 609.5D x 609.5H (mm)
Exterior: 58.8W x 37.6D x 81.3H (in)
  1493.6W x 955.1D x 2065.1H (mm)
Volume: 8 cubic ft
  226.53 liters
Temperature: -37°C to 180°C (-34.6°F to 356°F)
Rise Rate*: 5°C/min
Pull Down Rate*: 4°C/min
Electrical Supply:  208 VAC, 1 PH, 50HZ or 60HZ
Full Load Amps: 21 A
Facility Power Receptacle: NEMA L6-30
Refrigeration System:  Single stage
Mobility: Will fit through 48 in doorway
Approximate Product Weight: 680.0 lbs (308 kg) without added options
Approximate Shipping Weight: 790.0 lbs (358 kg) without added options

*Rise and pull-down rates calculated using IEC 60068-3-5 standard. Custom options available.

*Ideal ambient temperature conditions for test chambers is 23°C (73.4°F). The temperature and humidity range of assured performance is between 15°C and 25°C (59°F/77°F) and 30-50% relative humidity. Conditions outside this range may affect the performance of your test chamber and reduce the lifespan of the equipment.

Standard Features

STANDARD FEATURES

  • Temperature range: -37°C to 180°C (-34.6°F to +356°F)
  • Two SD-508-ATP
  • Mountable on a reinforced floor stand with locking casters. Includes t anchor brackets for mounting floor stand to facility floor.
  • External cabinet construction consists of heavy-gauge, cold-rolled steel with powder coat, bone color, and textured matte finish
  • Internal workspace constructed with type 304 stainless steel
  • Full opening door with adjustable manual latch
  • Interior light with exterior switch on cabinet face and operable on AESONE CONNECT
  • Port configuration: 2 rear wall access ports, and 2 with 4in (101.6mm) diameter for ATP Fixtures
  • Fully adjustable shelf, 1 wire-grid shelf, stainless steel
  • Forced air circulation
  • ATP mounting provisions, internal workspace
  • Multipane 12 in x 12 in heated viewing window>
  • Overall Dimensions on Floor Stand (in): 58.4W x 37.5D x 81.3H

Instrumentation

STANDARD INSTRUMENTATION

CONTROLLER   AESONE CONNECT allows for secure remote access to your test chamber and gives easy profile creation, operational control, and data management.

SD Series Test Chambers have incorporated AESONE CONNECT creating a robust connected device for a standard and uniform interface making test chamber operation easier.

AESONE XCHANGE is hardware internal to the chamber that increases your functionality

  • Connects the test chamber controller to your network for increased accessibility
  • On-board multi-year data logging

AESONE SOFTWARE gives you the power to

  • Remote chamber operation and monitoring, 
  • Web-based test profile creation, 
  • Viewing current data and historical data visually
  • Secure multi-level user access 

ADDITIONAL ONBOARD INSTRUMENTATION

WATLOW F4T PROGRAMMABLE CONTROLLER. Features a 4.3” capacitive touchscreen minimizing the chance for error. Ethernet communication supported by AESONE HARDWARE. View more controller options

SENSORS.

  • TEMPERATURE SENSOR. Measured at the point of airflow entering the workspace. Class A, 3-wire, 100 Ohm Platinum RTD sensor

OPTIONAL DATA LOGGERS. View additional data loggers

Safety Features

STANDARD SAFETY FEATURES

  • HIGH/LOW TEMPERATURE CONTROL. An adjustable temperature limit controller allows you to program the highest and lowest temperature of the device under test, protecting products from compromising temperature exposure.
  • Over Temperature Heat Fusing
  • Refrigeration system over-pressure safety circuit
  • TUV Rheinland certified meeting UL 61010-1, US, C, E

Optional Features

  • Custom Porting Sizes
  • Custom window size
  • Heater for Window
  • LN2 Boost Cooling
  • CO2 Boost Cooling
  • Water Cooled Condenser
  • Desiccant Air Dryer to ensure conditions as low as 5% RH 
  • Dry Nitrogen Purge to keep condensation and frost from accumulating on DUT surfaces at cold temperatures and to prevent oxidation of metals surfaces at high temperatures
  • Battery testing safety features