The BD-927 Model is a laboratory oven capable of providing highly-stable, controlled, forced hot air circulation. BD-927 Laboratory Ovens are standard laboratory ovens that provide dry, air-circulated heat as an environmental testing condition.
CONSTRUCTION. These laboratory ovens feature a durable powder-coat finish over heavy-gauge steel, with stainless steel internal construction, and highly-efficient, low K factor, thermal insulation, this is built to last for many years.
CIRCULATION SYSTEM. The ability to maintain precise temperatures occur with help from direct, drive-powered fans. The performance level operates the internal workspace temperature from +38°C* to +220°C (100.4°F to +428°F).
HEATING SYSTEM. The oven features an industry-leading heating system equipped with low-watt density, ceramic core, nichrome heaters for accurate temperatures within ±0.5°C, as measured at the sensor.
BD Series Laboratory Ovens are ideal for:
INSTRUMENTATION. The standard BD models have a manual-set digital controller. For more sophisticated test profiles, upgrade incorporates AESONE CONNECT creating a robust connected device giving you a standard and uniform interface that makes operation easier. Learn more in the Instrumentation Tab below.
With the popularity of rapid cooling, BD-927 Models have spawned a standard liquid-cooled model. Take a look at BD-100 Model Laboratory Ovens to incorporate liquid LN2 or CO2 cooling or cryogenic cooling into your laboratory oven.
Interior: | 36.0W x 36.0D x 36.0H (in) |
914.4W x 914.4D x 914.4H (mm) | |
Exterior: | 45.0W x 42.4D x 63.0H (in) |
1143W x 1077D x 1600.2H (mm) | |
Volume: | 27 cubic ft |
764.55 liters | |
Temperature: | 38°C to 220°C (100.4°F to 428°F) |
Rise Rate*: | 4.88 °C/min |
Electrical Supply: | 208 VAC, 1 PH, 60 HZ |
Full Load Amps: | 51 A |
Facility Power Receptacle: | Hardwired |
Mobility: | Will fit through 48 in doorway |
*Rise and pull-down rates calculated using IEC 60068-3-5 standard. Custom options available.
*Ideal ambient temperature conditions for test chambers is 23°C (73.4°F). The temperature and humidity range of assured performance is between 15°C and 25°C (59°F/77°F) and 30-50% relative humidity. Conditions outside this range may affect the performance of your test chamber and reduce the lifespan of the equipment.
Notice for California Customers (SB1206):
For test chambers shipped to California, please consult our sales team regarding the use of R449A refrigerant (where applicable). This substitution ensures compliance with SB1206 regulations.
Temperature is manually set with one Watlow EZ-ZONE PM controller (non-programmable)
BD Series Test Chambers can incorporate AESONE CONNECT creating a robust connected device for a standard and uniform interface making test chamber operation easier.
AESONE XCHANGE increases your functionality
AESONE SOFTWARE gives you the power to
WATLOW F4T PROGRAMMABLE CONTROLLER. Features a 4.3” capacitive touchscreen minimizing the chance for error. Ethernet communication supported by AESONE HARDWARE. View more controller options
HIGH/LOW TEMPERATURE CONTROL. An adjustable temperature limit controller allows you to program the highest lowest temperature the device under test, protecting products from compromising temperature exposure.